TB 9-6625-2798-50
(5) Push PRESS TO TEST button. Adjust
(7) Short COMMON and FLOOR AND
decade resistor until test instrument
EQUIPMENT terminals on the test
indicates value shown in column 1
instrument.
(8) Push PRESS TO TEST button on test
be between MIN and MAX values shown
instrument and zero the meter. Release
button.
decade resistor does not indicate between
(9) Remove
shorting
lead
from
test
instrument.
below.
(6) Remove test leads from the dec ade
resistor.
megohm and greater.
Table 4-1. Performance Check-Floor And Equipment (Less Than 1 Megohm)
TEST INSTRUMENT
INDICATION
RESISTOR IN OHMS
0
MINIMUM
MAXIMUM
10K
0
50
50K
9,800
10,200
100K
49,000
51,000
200K
98,000
102,000
500K
196,000
204,000
1 MEG
490,000
510,000
980,000
Figure 4-2. Performance check setup for 1 megohm and greater
(floor and equipment)
5