TB 9-6625-2798-50
(5) Push PRESS TO TEST button. Adjust
4-3. Final Procedure.
decade resistor until test instrument
indicates value shown in column 1
a. Disconnect leads and adapters and secure test
equipment. Place cover on test instrument.
be greater than minimum value shown in
b. In accordance with TM 38-750, annotate and
Release
button.
When the test instrument does not perform
within tolerance, annotate and affix DA Form
b. Adjustments. No adjustments can be made.
2417 (Unserviceable or Limited Use) tag.
Table 4-3. Performance Check-Personnel
TEST INSTRUMENT
INDICATION
DECADE RESISTOR IN
DECADE RESISTOR IN
OHMS
OHMS
MINIMUM
MAXIMUM
1 MEG
980,000
1,020,000
By Order of the Secretary of the Army:
E. C. MEYER
General, United States Army
Chief of Staff
Official:
J. C. PENNINGTON
Major General, United States Army
The Adjutant General
Distribution:
To be distributed in accordance with DA Form 12-34A, requirements for Section II, Calibration Procedures Publication.
U.S. GOVERNMENT PRINTING OFFICE : 1980--765016/1059
8