TB 9-6625-340-35
10. Attenuation
NOTE
This paragraph may be omitted for all systems codes except
FOO.
a. Performance Check
(1) Connect attenuation measurement equipment as shown in figure 4.
Figure 4. Attenuation measurement - equipment setup.
(2) Select one TI attenuator.
will be within 0.3 dB of attenuation value specified on calibration chart.
(4) Repeat technique of (2) and (3) above at remaining frequencies indicated on
calibration chart of TI attenuator.
(5) Repeat (2) through (4) above for remaining TI attenuators.
b. Adjustments. No
adjustments can be
made. If
necessary, correct calibration chart
or prepare new calibration chart.
10