Input Zero Set
Figure 6. Attenuator compensation - equipment setup.

Calibration Procedure For Electronic Counter, Model 5245L And 5245M, And Internal Timer,, Model 5262A Page Navigation
  6    7    8    9    10  11  12    13  
img
. . . . . .
TB 9-6625-781-24
Figure 4. Interval timer - right view.
(3) Disconnect multimeter from STOP input and connect START input.
(4) Adjust A3R6 (fig. 5) on start channel board for zero ( 10 mV) indication
on multimeter.
Figure 5. Interval timer left view.
b. Adjustments. No further adjustments can be made.
17. Attenuator Compensation
a. Performance Check
(1) Connect equipment as shown in figure 6.
11