TB 9-6625-975-35
Figure 3. Adjustment locations.
15. Voltage Check
a. Performance Check
(1) Connect digital voltmeter (A1) positive dc terminal to GROUND terminal, and
common terminal to GUARD or SHIELD terminal, using leads and adapters (B3, B4, B1,
and B2).
(2) Set TI RANGE switch to 1000V MΩx2. Rotate generator crank cw at normal
speed. If digital voltmeter does not indicate 1000 V dc 25 V dc perform b(1) below.
10