TB 9-6670-248-50
SECTION II
5. Equipment Required. Table 1
lists minimum
use specifications of
equipment
required for calibration performance checks and adjustments. Table 1 is provided to assist
in the selection of required equipment. For specific item identification refer to section III.
Table 1. Minimum Specifications of Equipment Required
Item
Minimum use
number.
Common name
specifications
A1
WEIGHTS
RANGE: 0 to 110 lb
ACCURACY: 1.33 oz
A2
WEIGHT SET
RANGE: 0 to 110 lb
ACCURACY: 1.33 oz
6. Preliminary Procedure
NOTE
Personnel should familiarize themselves with the entire
a. Place the UUT on a clean flat level surface.
b. Check that
counterpoise weight hook (fig. 1) is
installed at
end of
beam (fig. 1) of
UUT.
NOTE
UUT type III style 3 does not have a counterpoise weight hook.
c. Check that trig
loop lock (fig. 1) is in the
unlocked position.
d. With the
poise (fig. 1) set on
zero, check that
beam is
balanced. If
necessary, rotate
zero balance control (fig. 1) clockwise or counterclockwise as required to balance the beam.
NOTE
When the UUT is not within tolerance, the deficiency must be
corrected before continuing with the procedure.
7. Beam and Counterpoise Weights
a. Performance Check
(1) Place 10-pound weight from weight set (A2, table 1) in center of platform of
UUT.
3