TB 9-6670-258-35
Paragraph
Page
System calibration alternate technique
(reference) ............................................................
Final procedure .....................................................
CALIBRATION PROCESS FOR REVERE,
MODELS C55800-3-50, 155800-00,
155800-01, AND 155800-03
Preliminary instructions ......................................
Equipment setup ...................................................
Load test ................................................................
Zero and phase check alternate technique
(reference) ............................................................
Load test alternate technique (reference)
Final procedure .....................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Aircraft Weighing Kit, BLH, Model C1; Evergreen, Model Roadrunner II; and
Revere, Models C7500, C46500, C55800-4-25, C55800-3-50, CS7, 155800-00, 155800-01,
and 155800-03. The manufacturers' manuals, TM 55-6670-200-14&P, and TM 55-6670-201-
13&P were used as the prime data sources in compiling these instructions. The equipment
being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 3
hours, using the physical technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Range: 0 to 50,000 lbs (3 cells)1
Force
Accuracy: 0.25% of reading or 25 lbs, whichever is greater
(25 lbs from 0 to 10,000 lbs)
1
Procedure limitation. Three load cells (50,000 lbs each) calibrated only from 0 to 15,000 lbs. Alternate technique
(reference) covers complete range of 50,000 lbs.