TB 9-6670-258-35
NOTE
Do not disconnect TI cells from TI indicator except as necessary
when making setups.
f. Connect cable to appropriate connectors on load cell No. 2 and load cell indicator.
g. Connect load cell indicator to a 115 V ac power source. Energize all equipment and
allow unit to warm-up for 15 minutes.
Figure 1. Aircraft weighing kit -
equipment setup.
h. Set TI CELL BALANCE switch to ON and CELL SELECTOR switch to 1.
i. Set TI WEIGHT dial and INITIAL LOAD switch to 0.
j.
Adjust TI ZERO SET control No. 1 for balance in no load condition.
k. Set TI INITIAL LOAD switch to 5000 (15,000 pounds, using load cell No. 3) and
WEIGHT dial to 0 pounds.
l.
Exercise load cell and TI cell as described in (1) through (5) below:
(1) Apply force slowly to load cell and TI cell to obtain approximately 5000 pounds
(15,000 pounds, load cell No. 3).
5