TB 9-6670-258-35
(7) After matching errors are obtained, apply loads to any cell in 3000 pound increments.
(8) Remove TI panel.
(9) Remove cover of board and component assembly (fig. 4).
(10) Adjust resistor SPOOL FINGERS (fig. 4) for best linearity. (R)
NOTE
When sliding these fingers, each cell error will shift a
similar value. If cell with highest error value at an
applied load is being adjusted to a lower value, cell with
lowest error value cannot be allowed to drop below the
tolerance limits.
(11) Replace board and component cover and TI panel.
(12) Repeat 8 a above.
8