TB 9-6670-258-35
(7) Connect all three TI cells to TI indicator using proper cables (color coded).
NOTE
Do not disconnect TI cells from TI indicator except as necessary
when making setups.
(8) Connect cable to appropriate connectors on load cell No. 4 and load cell indicator.
(9) Connect load cell indicator to a 115 V ac power source. Energize all equipment
and allow unit to warm-up for 15 minutes.
(10) Zero set load cell in no load condition.
(11) Set TI CELL BALANCE switch to ON and CELL SELECTOR switch to 1.
(12) Turn TI WEIGHT dial and INITIAL LOAD switch to 0.
(13) Adjust TI ZERO SET control No. 1 for balance in no load condition.
(14) Set TI INITIAL LOAD switch to 45 and WEIGHT dial to 5000 pounds.
11