TB 9-6685-330-35
NOTE
TI should be mounted in the vertical position for testing.
NOTE
TI should be on same level with the PNEUMATIC PRESSURE
STANDARD.
a. Fill in the
general data on the
top of the
Differential Pressure Gage Calibration
WARNING
To prevent injury to personnel or damage to equipment, make
certain that all components are within the range of the unit to
be calibrated and all connections are securely sealed prior to
applying pressure to TI. Never attempt to tighten connections
with pressure applied.
c. As
shown in
figure 2, connect the WATER PUMP NITROGEN REGULATOR to the
NITROGEN TANK. Then from the WATER PUMP NITROGEN REGULATOR, connect to
the first "T" FITTING. Then from the first "T" FITTING, connect to the INLINE VALVE.
Then from the INLINE VALVE, connect to the second "T" FITTING. Then from the second
"T" FITTING, connect to the 0-250 PSIA port of the PNEUMATIC PRESSURE
STANDARD.
e. Cap off the
open "T" FITTING
(fig. 2) prior to
connecting the TI.
f. Recheck all connections