TB 9-6685-330-35
SECTION I
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Pressure Gages Differential (General). Calibration and Repair Requirement
Worksheet, DA Form 3758, and various manufacturers' manuals were used as the prime
data sources in compiling these instructions. This bulletin complies with ANSI B40.1-1980
specification for indicating dial type elastic element pressure gages. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Figure 1 is a blank
calibration data sheet used as a
guide for keeping records, and
can be reproduced locally.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument
Performance
parameters
specifications
Pressure
Range: 0 to 60 psi
Accuracy: 5% of reading
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI.
2