TB 9-6685-363-50
(10) If TI
does not fall within the required tolerance, individual calibration of
the indicating dial should be
followed
as outlined in TB 9-6685-319-50 using the manufacturer's calibration record of cell output pressure in psi (fluid pressure
vs indicating dial reading in lb force) for each individual TI load cell. See figure 4 (p 16). Approximately 40 psi (2.80
2
2
kg/cm ) will be zero on dial and 1000 psi (70.00 kg/cm ) will be 20,000 pounds (lb) (9072.00 kg) force on dial.