(11) If the TI indicating dial is within tolerance, the TI manufacturer should be consulted for factory check of
probable defective load cell.
9. Final Procedure.
a. Deenergize and disconnect equipment.
b. In accordance with TM 38-750, annotate and affix DA Label 80 (US Army Calibrated Instrument). When the TI
receives limited or special calibration, annotate and affix DA Label 163 (US Army Limited or Special Calibration). When
the TI cannot be adjusted within tolerence, annotate and affix DA Form 2417 (US Army Calibration System Rejected
Instrument).