TB 9-6625-242-24
NOTE
Tap face of TI lightly with finger while zero checking and
before making a reading.
NOTE
Turn zero-adjusting screw in direction that will return pointer
to zero through least degree of travel.
f. Connect load cell indicator to 115 V ac power source. Energize equipment and allow
unit to warm up for 30 minutes.
g. Position controls on load cell indicator as appropriate for load cell to be utilized
with TI.
h. Exercise load cell and TI as described in (1) through (5) below:
Figure 1. Dynamometer accuracy -
equipment setup.
CAUTION
Do not exceed capacity load on either TI or bad cell.
(1) Apply tension (force) slowly to load cell and TI to obtain an approximate
capacity load.
(2) Release tension on load cell and TI to obtain an approximate zero indication on
load cell indicator.
(3) Repeat (1) and (2) above two more times.
(4) Verify that TI indicates zero.
4