TB 9-6695-301-50
(1) Press the F4 key on the TI indicator until you see the UNIT soft key. Press the
UNIT soft key as necessary to select mV/V units to be displayed. Press the SET key then
press the F3 key to select the Amplifier menu. Press the SIGNAL
and
keys to select Display from the Amplifier menu. Press the SIGNAL
and
keys to key-pad set the decimal places to 6.
(2) Zero the load cell indicator.
(3) Apply force for load cell indicator indications listed in the Applied Force column
of the Transfer Load Cell Calibration Spreadsheet (TLCCS.xls) for the TI load cell and
direction being calibrated. Enter TI indicator readings in the Transfer Load Cell
Calibration Spreadsheet (TLCCS.xls) for the TI load cell being calibrated.
(4) Rotate TI load cell 120.
(5) Repeat (1) and (4) above for a total of three runs.
(6) Repeat b(1) thru (5) above for tension. If the TLCCS.xls indicates a passing
condition (after all entries are made) then continue to next step. If the TLCCS.xls indicates
a failing condition then TI load cell is bad.
(7) Press the SET button and then press the F3 key to select the Amplifier menu.
key to select Transducer from Amplifier menu.
(8) Press the
keys to ensure that the Transducer
(9) Press the SIGNAL
and
menu is set according to (a) thru (d) below.
(a)
Type:
SG full bridge.
(b)
Excitation:
5V
(c)
Unit:
lb
mV/V
(d)
Zero point:
0.000
0.00000
(e)
Nom.-value:
(TI nominal full scale)
3.00000
(f)
K-Factor:
0.0000
keys to select Ext. fct....
(10) Press the SIGNAL
and
keys and key-pad input the C0-C3
(11) Press the SIGNAL
and
coefficients from the TLCCS.xls.
keys to select OK.
(12) Press the SIGNAL
and
keys to select Adjust amplifier. Press
(13) Press the SIGNAL
and
the ESC key.
(14) Press the F3 key to select the Amplifier menu. Press the SIGNAL
keys to select Display from the Amplifier menu. Press the SIGNAL
and
and
keys and key-pad set the decimal places to 2. Press the ESC key.
(15) Press the SET key. On the Save set-up? menu select Yes by pressing the
key.
(16) Repeat 10a. above; if indications are not within tolerance, then TI load cell is
bad.
P
IN NO: 080761-001