TB 9-4910-556-35
(2) Set calibrator for 10.000 V ac and 100 Hz.
(3) Set calibrator to OPERATE mode.
(4) Press and release STEP key on SC. SC will display:
AC VOLTAGE
Range: 30 mV to 140 V rms
Then: Between 9.42 and 10.58 V rms.
(5) Press and release STEP key on SC. SC will display:
Range: 40 to 1000 Hz
Then: Between +99.0 and 101.0 Hz.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect cable W2 red and black leads to resistance standard.
(3) Press and release STEP key on SC. SC will display:
RESISTANCE
Range: 0 to 1000 kΩ.
Then: Between 1.372 and 1.428 kΩ.
b. Adjustments. No adjustments can be made.
a. Performance Check