TB 9-5855-1892-24
(4) Connect remaining end of ac power cable to 115 V ac source.
(5) Set ON/OFF switch to ON.
NOTE
READY lamp will initially illuminate then extinguish for
approximately 10 seconds. READY lamp will then blink once
a second.
(6) Verify READY lamp blinks approximately once a second and POWER lamp
illuminates; if not, perform 8 b(1) below.
g. Set ON/OFF switch to OFF.
h. Perform steps (1) through (7) below:
(1) Rotate target select knob to MRT.
(2) Set ON/OFF switch to ON.
NOTE
READY lamp will initially illuminate, then extinguish for
approximately 10 seconds. READY lamp will then blink once
a second.
(3) Verify READY lamp blinks approximately once a second; if not, perform 8 b (1) below.
(4) Set TI ON/OFF switch to OFF.
(7) Loosely replace large cover plate (fig. 2).
a. Set target select knob to BAR.
b. Connect equipment as shown in figure 5 and as described in (1) through (5) below:
NOTE
Equipment used to mount thermocouples to TI may vary at
each lab. The following instructions describe the most common
equipment in use.
(1) Install thermocouple mounting post (pair) (fig. 5) on base of TI (approximately
center of second slot pair from exit port).
(2) Install thermocouple harness wire to thermocouples (connectors are keyed and labeled).
(3) Clamp differential thermocouple set (pair) in laboratory clamps and install on
thermocouple mounting posts, using mounting post clamps (fig. 5).
(4) Using the wood end of a cotton tip applicator, daub a small amount of thermal
conducive paste on bottom of HEATER PLATE behind BAR TARGET (fig. 6).