TB 9-6625-1982-24
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
DC508OPT01 AND DC508AOPT01
33. Preliminary Instructions
a. The instructions outlined in paragraphs 33 and 34 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance checks where applicable.
CAUTION
To avoid damage to the plug-in circuitry, power must be OFF
on the mainframe power module before connecting the TI.
a. Remove left or right side covers of TI as necessary to make connections or
adjustments. Replace cover(s) after completing adjustments.
b. With power OFF, connect TI to mainframe power module using extender.
c. Position controls as listed in (1) through (7) below:
(1) STANDBY pushbutton in.
(2) DIRECT RESOLUTION switch to 1 Hz.
(3) DISPLAY TIME variable control fully ccw.
(4) RESOLN MULT pushbutton to X1 (out).