TB 9-6625-2071-24
indicates -2.000 V dc 0.002 V dc.
(19) Adjust oscilloscope controls as necessary to view waveform. Adjust A11R26 (fig. 1)
(20) Move oscilloscope Vertical 1 connection from TI CHANNEL A MARKER
OUTPUT to TI CHANNEL B MARKER OUTPUT.
(21) Adjust oscilloscope controls as necessary to view waveform. Adjust A11R17 (fig. 1)
(22) Set function generator Offset to 0 V dc.
a. Deenergize and disconnect all equipment and reinstall TI protective cover.
b. Annotate and affix DA label/form in accordance with TB 750-25.