A TRIGGER
A TRIGGER LEVEL
adjustments
stable display test
MODE
SLOPE
b
YES
NO
b
NORM
IN:
b
P-P AUTO
IN:
b
P-P AUTO
OUT:
(30) Set HORIZONTAL MODE switch to B adjust B INTENSITY control for
suitable viewing.
(31) Verify a stable display can be obtained by adjusting B TRIGGER LEVEL
control in a position other than B RUNS AFTER DLY; if not, perform b below.
(32) Press B TRIGGER SLOPE pushbutton to IN and verify a stable display can be
obtained by adjusting B TRIGGER LEVEL control in a position other than B RUNS
AFTER DLY; if not, perform b below.
(33) Position controls as listed in (a) through (d) below:
(a) VERTICAL MODE CH 1 BOTH CH 2 switch to CH 1.
(b) HORIZONTAL MODE switch to A.
(c) A AND B SEC/DIV switches to .05 Ps.
(d) B TRIGGER SLOPE pushbutton to OUT.
(34) Connect oscilloscope calibrator CHAN 1 to TI CH 1 using a 50 :
feedthrough termination.
(35) Set oscilloscope calibrator LEVEL SINE output to 100 MHz and approximately
100 mVpp for 1.5 divisions of vertical display on TI.
(36) Set A TRIGGER pushbutton to first row listed in table 21 and adjust A
TRIGGER LEVEL control to obtain a stable display. If a stable display cannot be
obtained, perform adjustments listed in table 21.
(37) Repeat technique of step (36) above for remaining A TRIGGER pushbutton
combinations listed in table 21. If a stable display cannot be obtained for each combination,
Test instrument
Test instrument
Test instrument
A TRIGGER
A TRIGGER LEVEL
adjustments
stable display test
MODE
SLOPE
b
Yes
No
b
NORM
IN:
b
P-P AUTO
IN:
b
P-P AUTO
OUT:
29