TB 9-6625-2247-40
29.00
23.11
33.00
20.15
35.00
18.67
37.00
17.18
b. Adjustments. No further adjustments can be made.
12. Automatic Amplitude Discrimination Test
a. Performance Check
(1) Connect equipment as shown in figure 6, connection A.
for a -25 dBm indication on power meter.
(3) Connect equipment as shown in figure 6, connection B.
output controls for a -5 dBm indication on power meter.
(5) Disconnect 50 Ω termination from power splitter, POINT A (fig. 6, connection B)
and connect signal generator cable to power splitter, POINT A (fig. 6, connection B).
(6) Disconnect power meter from power splitter, POINT B (fig. 6, connection B) and
connect TI to power splitter, point B (fig. 6, connection B). TI will indicate 18 GHz 4 Hz.
(7) Connect equipment as shown in figure 6, connection A.
controls for a -11 dBm indication on power meter.
(9) Connect equipment as shown in figure 6, connection B.
controls for a -5 dBm indication on power meter.
(11) Disconnect 50 Ω termination from power splitter POINT A (fig. 6, connection B)
and connect signal generator to power splitter POINT A (fig. 6, connection B).
(12) Disconnect power meter from power splitter, POINT B (fig. 6) and connect TI to
power splitter, POINT B (fig. 6). TI will indicate 2.5 GHz 4 Hz.