TB 9-6625-2267-35
(5) Set calibrator for a -190 mV dc output. Adjust C30 (fig. 1) until TI indication is
within 1 digit of indication recorded in (4) above (R).
(6) Repeat (1) through (5) above until no further adjustments are required.
(7) Press RANGE 2 pushbutton.
(8) Set calibrator for a 1.9 V dc output. Adjust R57 (fig. 1) until TI indicates
+1.9000 1 digit (R).
(9) Set calibrator for a -1.9 V dc output. Adjust R57 (fig. 1) until TI indicates
-1.9000 1 digit.
(10) Repeat (8) and (9) above until no further adjustments are required.
(11) Press RANGE 20 pushbutton.
(12) Set calibrator for a 19 V dc output.
+19.000 1 digit (R).
(13) Set calibrator for a -19 V dc output. Adjust R4 (fig. 1) until TI indicates -
19.000 1 digit.
(14) Repeat (12) and (13) above until no further adjustments are required.
(15) Press RANGE 200 pushbutton.
(16) Set calibrator for a 190 V dc output. Adjust R6 (fig. 1) until TI indicates
+190.00 1 digit (R).
(17) Press RANGE 1200V pushbutton.
(18) Set calibrator for a 1000 V dc output. Adjust R8 (fig. 1) until TI indicates
+1000.0 1 digit (R).
6