TB 9-6625-2296-24
(g) DATA ENTRY 3 0 M ..............
30 MHz
(h) Ret (F5) ....................................
Return
(4) Set measuring receiver to measure RF power at 30 MHz.
(5) Set measuring receiver to ratio mode.
(6) Press TI DATA SCROLL arrow keys to manually sweep TI frequency from
500 kHz to 399 MHz entering TI display frequency into measuring receiver. If measuring
receiver indication does not remain within 2.0 dB, perform b below.
(7) Press TI DATA SCROLL arrow keys to manually sweep TI frequency from
400 MHz to 999 MHz entering TI display frequency into measuring receiver. If measuring
receiver indication does not remain within 3.5 dB, perform b below.
(8) Set TI frequency to 30 MHz. Set measuring receiver to measure tuned RF level
and manually enter 30 MHz. Perform measuring receiver calibration.
NOTE
(9) Press TI Level (F3) (dBm display) key and adjust output level for the following
dBm indications: 0, -1, -2, -4, -8, -16, -32 and -64. Measuring receiver indications will be
within 2.0 dBm of TI indications.
(10) Adjust output level from -90 dBm to -110 dBm in 10 dB steps. Measuring
receiver indications will be within 2.5 dBm of TI indications.
b. Adjustments
(1) Press key sequence listed in (a) through (l) below:
(a) MODE MTRS ..........................
Meter Menu
(b) AUX (F6) ..................................
Auxiliary Functions Menu
(c) MEMORY RCL .......................
Recall
(d) FIELD SELECT ..................
Highlight 10. Factory Defaults
(e) DATA ENTRY ENTER ...........
Factory Defaults recall
NOTE
Assure highlighted area indicates YES before proceeding.
(f) DATA ENTRY ENTER ...........
Factory Defaults restored
(g) MODE RF GEN ......................
RF Gen Display
(h) SETUP .....................................
Gen Menu
(i) DATA ENTRY 5 ......................
RF Gen Setup
DATA ENTRY 2 ......................
(j)
RF Gen Level
(k) DATA ENTRY 0 ENTER .......
0 dBm
28