TB 9-6625-2316-35
5. Accessories Required. The
accessories required for the
calibration are common usage
accessories, issued as indicated in paragraph 4 above, and are not listed in this calibration
procedure. The following peculiar accessories are also required for this calibration:
Extender Card, part number 385E302; Electric Connector Assembly, part number
M55074/4 (NSN 5935-01-142-9742).
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
(part number)
Common name
Minimum use specifications
Range: Dc voltage: (30 mV to 3.0 V)
John Fluke, Model 8506A/CT
(p/o MIS-35947)
Accuracy: 0.5%
Range : Observe 1000 Hz signal
Tektronix, Type 2465BOPT46
Accuracy: 0.5%
(2465BOPT46)
Biddle-Gray, Model 71-650
Range: 16.78 kΩ
(71-650)
STANDARD NO. 1
Accuracy: 0.5%
Biddle-Gray, Model 71-631
Range: 19.6Ω to 700 kΩ
(7910328)
STANDARD NO. 2
Accuracy: 0.5%
Beckman, Model CR10M
Range: 9 and 10 MΩ
(8598965)
STANDARD NO. 3
Accuracy: 0.75%
SECTION III
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in the TM 11-6625-3292-40 and the
manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
3