TB 9-6625-2341-35
(b)
Set oscilloscope calibrator to Output voltage settings at 1 kHz while
adjusting edit amplitude output dial for 4 vertical divisions.
NOTE
If necessary, adjust TI Trigger level by toggling ▲ or ▼
pushbutton for a stable display.
NOTE
If necessary, toggle left side ▲ MOVE ▼ pushbutton center to
TI display.
(c)
If oscilloscope calibrator Voltage error is not within specified limits,
perform b below.
(d)
Set oscilloscope calibrator to standby.
Test instrument
Oscilloscope calibrator
Range settings
Output volt
Voltage error
(mV/Div)
settings at 1 kHz
(%)
(V p-p)
Min
Max
5
.02
-3
3
10
.04
-3
3
20
.08
-3
3
50
.2
-3
3
(7)
(8)
(F1) CLOSE which should display/highlight CONTR.
OSCILLOSCOPE CALIBRATOR
(9)
Press pushbutton sequence as listed in (a) through (e) below:
(a)
"Input A" to off by pressing (1) through (7) below:
(1)
INPUT A.
(2)
(F1) MORE INPUT A.
(3)
Toggle ▲ or ▼ to display/highlight INPUT A.
7