TB 9-6625-2341-35
a. Performance Check
(1) Reset by pressing pushbuttons sequence (a) through (i) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle ▲ or ▼ to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle ▲ or ▼ to display/highlight RESET METER.
(f) (F5) SELECT ITEM.
(g) All meter settings have been set to original setting then (F1) CLOSE which
should display/highlight RESET METER.
(h) (F1) CLOSE should display/highlight RESET MENU.
(i) (F1) CLOSE.
(2) RESET calibrator.
(3) Set measure to VAC by pressing (a) through (d) below:
(a) MEASURE MENU to display/highlight (V, Hz, dB, TIME).
(b) (F2) V.
(c) Toggle ▲ or ▼ to display/highlight Vrms AC METER.
(d) (F5) SELECT ITEM.
(4) Set "Input A" probe to 1:1 by pressing (a) through (g) below:
(a) INPUT A.
(b) (F5) PROBE A MENU.
(c) Toggle ▲ or ▼ to display/highlight PROBE on INPUT A.
(d) (F5) SELECT ITEM which should display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
(e) ▲ or ▼ to display/highlight 1:1.
(f) (F5) SELECT ITEM.
(g) (F1) CLOSE which should display/highlight CONTR.
(5) Select display change by pressing (a) through (h) below:
(a) SUB MENU.
(b) (F1) MORE METER.
(c) Toggle ▲ or ▼ to display/highlight READINGS on DISPLAY.
(d) (F5) SELECT ITEM which should display/highlight (1, 2, 3, 4).
(e) Toggle ▲ or ▼ to display/highlight 1.
(f) (F5) SELECT ITEM.
(g) (F1) CLOSE.
(h) SCOPEMETER/MAIN MENU to display SCOPE/METER/Ω/EXT.mV
and etc.
(6) Perform steps as listed in (a) through (c) below for each row in table 6.
(a) Under INPUT A pushbutton left side, toggle (mV RANGE V) to display TI
Range settings.
(b) Set calibrator to Voltage output settings at selected Frequency.
(c) If TI Voltage indications are not within specified limits, perform b below.
(d) Set calibrator to standby.