TB 9-6625-2345-35
(4) Press signal generator #1 INSTRUMENT PRESET key.
dBm.
(6) Press TI keys as listed in (a) through (f) below:
(a) MARKER MENU, 5.
(b) 8, 0, kHz dBX.
(c) BLUE SHIFT.
(d) FREQ, 5, 0, 0, MHz +dBX.
(e) BLUE SHIFT.
(f) SPAN/DIV, 5, 0, Hz dB.
(7) Press TI SINGLE SWEEP key twice.
(8) When sweep is completed, press TI GREEN SHIFT then PEAK FIND keys.
(9) Marker frequency displayed on TI CRT will be within limits specified in first row
(10)Repeat technique of (5), (6) (c), (6) (d) then (7) and (8) above for remaining signal
generator #1 and TI settings in table 3. Marker frequency displayed on TI CRT will be
Signal generator #1
Test Instrument
Frequency
Amplitude
Frequency
Span/Div
Marker Freq. Display (MHz)
(dBm)
Min
Max
500
MHz
0
500
MHz
50 Hz
499.999865
500.000135
1.00 GHz
0
1.00 GHz
50 Hz
999.999756
1000.000235
1.50 GHz
0
1.50 GHz
50 Hz
1499.999665
1500.000335
(11) Press TI keys as listed in (a) through (g) below:
(a) BLUE SHIFT.
(b) FREQ, 5, 0, 0, MHz +dBX.
(c) BLUE SHIFT.
(d) SPAN/DIV, 1, MHz +dBX.
(e) COUNT (illuminated).
(f) BLUE SHIFT.
(g) COUNT RESOLN, 1, Hz dB.
dBm.
(13) Press TI SINGLE SWEEP key twice.
(14) Marker frequency displayed on TI CRT will be within limits specified in first row
9