SECTION V. CALIBRATION PROCESS FOR BALANCER ANALYZER CHADWICK-HELMUTH, MODELS 8500C AND 8500C+
Balance Test

Claibration Procedure For Model B4591, 11, 11A, 177M-6A, 135M-11, 8500C And 8500C+ Page Navigation
  7    8    9    10    11  12  13    14    15    16    17  
img
. . . .
TB 9-4920-463-24
d. Connect equipment as shown in figure 3 below.
Figure 3. Equipment connection.
e. Adjust gap between magnetic pickup and screw to 0.04 inch
NOTE
The following steps erase all TI memory.
f. Hold down TI START key and turn on power supply.
g. Wait until "Power-Up Self-Test" display appears, then release TI START key.
18. Balance Test
a. Performance Check
NOTE
These tests build successively.  Any settings not mentioned
must be left as they were in the previous test.  The use of
brackets [ ] denotes a soft key. To undo an entry, press TI
[RESTORE] soft key.
(1) Ensure only one No. 6 screw is installed above "h" on logo on Vibrex tester 11A
rotor disk.