Figure 7. Overlay sample.
Disk Drive Test

Claibration Procedure For Model B4591, 11, 11A, 177M-6A, 135M-11, 8500C And 8500C+ Page Navigation
  13    14    15    16    17  18  19  
img
.
TB 9-4920-463-24
Figure 8. FasTrak equipment connection.
(e) STATUS.
(8) After a few seconds, TI display will indicate approximately 900 RPM.
NOTE
If you cannot adjust FasTrak distance or alignment to obtain a
900 RPM indication on TI, there may be a problem with
FasTrak.
(9) Use a white card or piece of paper to cover half of Vibrex tester 11A rotating
disk. Place card close to, but not touching disk, to cover half of magnetic pickup sensor and
one vibration sensor.
(10) TI RPM indication will drop by approximately half.
(11) Turn off all lights in room or enclose setup in a darkened enclosure. TI RPM
will drop to zero when area is dark enough.