SECTION III. CALIBRATION PROCESS
Table 3. Wireless Diagnostic Sensor-Analog (WDS-A) Accuracy.

Calibration Procedure For Test Adapter Set, Smart Wireless Internal Combustion Engine, 13666129 And 13675052 Page Navigation
  1    2    3  4  5    6    7  
img
.
TB 9-4920-468-24
d. Set multimeter with wireless adapter (WLDMM) attached to ACV position.
Figure 1. SWICE Devices Connection Diagram.
e. Launch current SwiceCal program (see TB 43-180) from instrument controller.
NOTE
Spiral 2 is identified by a green case and green wireless
devices.  Spiral 3 is identified by a black case and black
wireless devices.
f. From set-up tab (Kit Selection), select SPIRAL 2 or SPIRAL 3 as applicable.
g. From set-up tab (Calibrator), set calibrator address. Press OK to return to main
screen.
h. Ensure that all tests are selected (SwiceCal default).
i. Press START TEST. SwiceCal will initialize and establish communication with all
wireless devices and automatically begin SELF TEST.
NOTE
Refer to IETM 9-6625-3637-23&P for troubleshooting if
SwiceCal fails to initialize, establish communication with all
wireless devices, fails SELF TEST or fails any other test.