TB 9-4931-393-35
e. Short POWER METER terminals (rear panel) using lead (B2). Leave short in
f. Set MOUNT RESISTANCE switch to 100Ω and FUNCTION switch to calibrate.
Allow 30 minutes for equipment to warm up and stabilize.
a. Performance Check
panel of TI, using cable assembly (B1).
(2) Set RANGE (MW) switch to .01.
(3) If dc voltmeter does not indicate between 632.14 and 632.78 mV (millivolts),
perform b below. Record indication.
and return to 115 V. The dc voltmeter indication will remain within limits specified in 3
above.
(5) Repeat technique of (2) and (3) above using settings, indications and adjustments
Test
Dc voltmeter indication
Adjustment dc
instrument
(V dc)
voltmeter
RANGE (MW)
Adjustments
indications
switch positions
Min
Max
(V dc)
.03
1.09485
1.09595
A1 R29 (R)
1.09540
.1
1.99900
2.00100
A1 R32 (R)
2.00000
.3
3.46246
3.46594
A1 R33 (R)
3.46420
1
0.632143
0.632777
A1 R36 (R)
0.632460
3
1.09489
1.09599
A1 R37 (R)
1.09544
10
1.99900
2.00100
A1 R40 (R)
2.00000
8
1.78796
1.78976
A1 R43 (R)
1.78886
6
1.54842
1.54998
A1 R46 (R)
1.54920
4
1.26428
1.26556
A1 R48 (R)
1.26492
2
0.893972
0.894868
A1 R51 (R)
0.894420
.01 1
0.446986
0.447434
---
---
10 1
1.41349
1.41491
---
---
MOUNT RESISTANCE switch to 200 Ω. These two checks are for low and high ends of RANGE switch for
1Set
200 Ω position of MOUNT RESISTANCE switch. If these checks are not within tolerance, it will be necessary to perform
RANGE switch adjustments, it may be necessary to readjust these several times to obtain best indication in all positions.
CHANGE 1