TB 9-4931-491-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Sweep width accuracy
200 kHz to 500 MHz, calibrated in 10 MHz
intervals
10 MHz
Band 1:
20 MHz
Band 2:
20 MHz
Band 3:
RF output accuracy
+10 to -80 dBm
0.5 dB to 500 MHz
0.75 dB from 1 to 500 MHz (when read with
Flatness at 10 dBm
measuring receiver)
1This
procedure covers only band 1 performance specifications check.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
Semiconductor Device (Coaxial Crystal Detector), Hewlett-Packard, Model 423A003
(7923182).
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
MEASURING RECEIVER
Flatness measurement:
Measuring Receiver system N5530S
consisting of: Spectrum Analyzer
Accuracy: 0.188 dB
Agilent, Model E4440A (E4440A),
Power Meter Agilent, Model E4419B
Frequency measurement:
(E4419B), and Sensor module,
Range: 50 to 500 MHz
Accuracy: 0.5%
Agilent Model N5532A opt. 504 (504)
Power measurement:
Range: +10 dB to -80 dB
Accuracy: 0.125 dB
Range: 1 to 50 V dc
Fluke, Model 8840A/AF05
Accuracy: 0.1%
(AN/GSM-64D)
Range: 1 mV/cm sensitivity
(OS-303/G)
Accuracy: 3%
Range: 10 to 110 MHz
Aeroflex, Model 2023B (2023B) or
Display Accuracy: 0.00125%
SG-1207/U (SG-1207/U)