SECTION III. CALIBRATION PROCESS
Figure 2. RF detector display.

Calibration Procedure For Sweep/Signal Generator Wavetek, Model 2001 And Sg-677 A/U Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . .
TB 9-4931-491-24
(10) Set SWEEP WIDTH to 520 MHz.
(11) POWER switch to on.
d. Allow 15 minutes for TI to warm up and stabilize.
8. Marker System
a. Performance Check
(1) Connect equipment as shown in figure 1.
Figure 1. Test equipment Marker System setup.
(2) Set oscilloscope for an X verses Y function operation.
(3) Adjust TI MARKER SIZE and oscilloscope vertical and horizontal controls to
obtain a pattern as shown in figure 2.
5