TB 9-4931-491-24
Paragraph
Page
Dial accuracy ........................................................
RF output level and vernier accuracy .................
RF output flatness................................................
Attenuator accuracy .............................................
Power supply ........................................................
Sweep rate voltage ...............................................
Source relationship ..............................................
Sweep drive voltage .............................................
Final procedure ....................................................
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Sweep/Signal Generator, Wavetek, Model 2001 and SG-677A/U. The
manufacturer's manual and TM 11-6625-2955-14&P were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 2 1/2
hours, using the dc and low frequency technique.
2. Forms, Records and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Frequency range
1 to 1400 MHz
1 to 500 MHz1
3 overlapping bands
Band 1:
Band 2:
450 to 950 MHz
Band 3:
900 to 1400 MHz
Start/stop, f, and cw
Operating modes
Frequency dial
10 MHz intervals
calibration accuracy
Band 1:
10 MHz
Band 2:
Band 3:
Footnote at end of table.