Figure 7. M1H balance adjustment.
Sweep Drive Voltage

Calibration Procedure For Sweep/Signal Generator Wavetek, Model 2001 And Sg-677 A/U Page Navigation
  11    12    13    14    15  16  17    18  
img
. . . . . . .
TB 9-4931-491-24
Figure 8. Sweep ramp.
(7) While observing waveshape in figure 8, adjust WAIT ADJ (fig. 3) for a wait time
of 1 ms (R).
16. Source Relationship
NOTE
The frequency accuracy of the TI is dependent on the +16 V
reference supply, the -16 V reference supply, the 32 V p-p
ramp, and the inverted 32 V sweep ramp. These four voltages
must be precisely adjusted in relation to each other to maintain
dial and display accuracy. (See figure 9.)
Figure 9. Relationship between sources.
16