Figure 5. M2H module.
Figure 8. Sweep ramp.

Calibration Procedure For Sweep/Signal Generator Wavetek, Model 2001 And Sg-677 A/U Page Navigation
  10    11    12    13    14  15  16    17    18  
img
. . . . . . . .
TB 9-4931-491-24
(4) If oscilloscope does not indicate a waveshape symmetrical about 0 V and 32 V p-p,
perform b below.
b. Adjustments
(1) Adjust M1H CENT control until waveshape is symmetrical about 0 V (fig. 3) (R).
(2) Adjust M1H SIZE control until amplitude is 32 V p-p (fig. 3) (R).
NOTE
These are preliminary adjustments only. Final adjustments
will be made in paragraph 16 below.
(3) Adjust VAR/MANUAL control fully ccw.
(4) If sweep time as displayed on oscilloscope is not 0.12 seconds (fig. 7), adjust M1H
INT BAL (fig. 3) (R).
Figure 7. M1H balance adjustment.
(5) Set SWEEP TIME SEC to LINE.
(6) If signal is not clamped at -16 V (fig. 8), adjust M1H CLAMP (fig. 3) (R).