Figure 7. Models 86210A and 86220A-H80 - bottom view.
Table 11. Power Output (Models 86200 Series)

Calibration Procedure For Sweep Oscillator Hewlett-Packard, Models 8620A And 86920C With Rf Plug-Ins, Models 86200 And 86300 Series; And Modules 8621A And 8621B Page Navigation
  20    21    22    23    24  25  26    27    28    29    30  
img
. . . . . .
TB 9-4931-495-24
Figure 9. Model 86260A-H80 - top view.
Table 9. Frequency Stability Specifications
Test instrument
Voltage change from
10 dB power
Model
105 to 125 V Ac
level change
20 kHz
20 kHz
86210A
20 kHz
20 kHz
86220A-H80
20 kHz
1 MHz
86230B-H80
30 kHz
1 MHz
86241A-H80
40 kHz
4 MHz
86242A-H80
40 kHz
2 MHz
86250B-H80
180 kHz
6 MHz
86260A-H80
Table 10. Manual Sweep (Model 86200 Series)
Test instrument MANUAL
Test Instrument MANUAL
Test
control fully ccw
control fully cw
instrument
microwave frequency counter
microwave frequency counter
models
indications (GHz)
indications (GHz)
Min
Max
Min
Max
86210A
---
<0.010
0.340
0.360
1
86220A-H80
---
<0.020
1.285
1.315
86230B-H80
1.785
1.815
4.185
4.215
86241A-H80
3.177
3.233
6.477
6.533
86242A-H80
5.860
5.940
8.960
9.040
86250B-H80
7.950
8.050
12.350
12.450
86260A-H80
12.330
12.470
17.930
18.070
Switch CONNECTION A from MICROWAVE FREQUENCY COUNTER INPUT 2 to INPUT 1.
1
26. Power Output
a. Performance Check
(1) Connect equipment as shown in figure 6, connection B.
(2) Position controls as listed in (a) through (c) below: