TB 9-4931-495-24
(9) Change output attenuator (fixed) in 1 step (10 dB) meter range control in 10 dB
increments to -30 dBm. Power meter will indicate within +2.0 dB of reference established
in (8) above.
b. Adjustments
(1) Model 86210A
(a) Set POWER LEVEL control to +13 dBm and adjust A1R14 (fig. 7) for
+13 dBm 0.2 dB power output indication (R).
(b) Set POWER LEVEL control to 0 dBm and adjust A1R13 (fig. 7) for 0
dBm 0.1 dB power output indication (R).
(2) Model 86220A-H80
(a) Set POWER LEVEL control to +10 dBm and adjust A1R1 (fig. 7) for +10
dBm 0.1 dB power output indication (R).
(b) Set POWER LEVEL control to 0 dBm and adjust A1R13 (fig. 7) for 0
dBm 0.1 dB power output indication (R).
(3) No adjustments can be made for models 86230B-H80, 86241A-H80, 86242A-
H80, 86250B-H80, and 86260A-H80.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
CALIBRATION PROCESS FOR RF PLUG-IN, MODELS 8621A AND 8621B WITH
RF PLUG-IN MODULE MODELS 86320A, 86330A, AND 86341B
a. Performance Check
(1) Connect equipment as shown in figure 6, connection A.
(2) Position controls as listed in (a) through (k) below:
(a) BAND selector to BAND 1 (0.1 to 2 GHz).
(b) START MARKER pointer to 0.1 GHz.
(c) STOP MARKER pointer to 2.0 GHz.
(d) MODE switch to MANUAL.
(e) TRIGGER switch to INT.
(f) TIME-SECONDS switch to 0.1-0.01.
(g) 1 kHz SW WV switch (rear panel) to OFF.