TB 9-4931-504-40
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Dual Trace Delayed Sweep Sampler, Tektronix Type 5S14N.
The
manufacturers' manual was used as the prime data source in compiling these instructions.
The above equipment will be referred to as the TI (test instrument) throughout this
bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
Performance specifications
VERTICAL SYSTEM
Risetime
350 ps or less, 10 to 90% of step pulse signal
Dc to 1 GHz
50 within 2%
2 mV/div to 0.5 V/div in 8 steps of 1, 2, 5 sequence
Deflection factor:
3% (with VARIABLE at CAL)
Accuracy
At least 2.5:1
Variable
Input signal range:
Maximum operation
2 V p-p (dc + peak ac) 2V
Dc offset range
2 V or more
Displayed noise(tangential)
2 mV or less, LOW NOISE switch "out"
Low noise operation
Displayed noise reduced by at least 5 times
Vertical signal out
0.2 V/div of deflection, 3%
Dot slash
Less than 0.1 div at 10 Hz and above
Delta t range
Shifts channel 2 at least 1 ns with respect to channel 1