TB 9-4931-535-24
(k) SWEEP ENTRY MANUAL.
(l) ENTRY STEP .
(5) Average TI marker amplitude indication will be <-80 dB.
key. Average TI marker amplitude indication will be <-80 dB.
(6) Press ENTRY STEP
key. Average TI marker amplitude indication will be <-80 dB.
(7) Press ENTRY STEP
key four times. Average TI marker amplitude indication
(8) Press ENTRY STEP
will be <-80 dB.
key. Average TI marker amplitude indication will be <-80 dB.
(9) Press ENTRY STEP
key. Average TI marker amplitude indication will be <-80 dB.
(10) Press ENTRY STEP
(11) Press keys and enter values using ENTRY keys as listed in (a) through (d) below:
ENTRY RECALL and then 1.
(a)
ENTRY CF STEP SIZE to 5 kHz.
(b)
SWEEP ENTRY MANUAL.
(c)
ENTRY STEP .
(d)
(12) Repeat (5) through (10) above.
(13) Press keys and enter values using ENTRY keys as listed in (a) through (d) below:
ENTRY RECALL and then 1.
(a)
ENTRY CF STEP SIZE to 100 kHz.
(b)
SWEEP ENTRY MANUAL.
(c)
ENTRY STEP .
(d)
(14) Repeat (5) through (10) above.
(15) Press keys and enter values using ENTRY keys as listed in (a) through (d) below:
ENTRY RECALL and then 1.
(a)
ENTRY CF STEP SIZE to 1 MHz.
(b)
SWEEP ENTRY MANUAL.
(c)
ENTRY STEP .
(d)
(16) Repeat (5) through (10) above.
input.
(17) Disconnect rear panel OVEN REF OUT from 50-75
(18) Reconnect rear panel OVEN REF OUT to EXT REF IN.
a. Performance Check
(1) Press keys and enter values using ENTRY keys as listed in (a) through (f) below:
ENTRY INSTR PRESET.
(a)
(b)
ENTRY CF STEP SIZE to 13.31 MHz.
(c)
RBW-VBW-ST ENTRY RES BW to 10 Hz.
(d)
INPUT ENTRY RANGE to -25 dBm using ENTRY STEP .
(e)