TB 9-4931-537-24
(2) Press multimeter FUNCTION/RANGE DCI pushbutton and set Guard switch
to To LO (depressed position).
(3) Preset (in standby mode) calibrator for a 100 A output. Press Currnt Output
soft-key until "Currnt Output NORMAL" is displayed.
(4) Set calibrator to Operate. Record multimeter indications in space provided.
(5) Set calibrator to remaining (m) A dc outputs listed in table 1.
Record
multimeter indications in space provided.
(6) Set calibrator to STANDBY, then press RESET pushbutton.
g. Ac Current. Calibrator, Fluke, Model 5720A/( ), Multimeter, Agilent, Model 3458A
and Amplifier, Fluke, Model 5725A/( ).
(1) Press multimeter FUNCTION/RANGE ACI pushbutton.
(2) Set calibrator to (, m) A, (k) Hz outputs listed in table 1. Record multimeter
indications in space provided.
(3) Set calibrator to STANDBY, then press RESET pushbutton.
h. Resistance. Calibrator, Fluke, Model 5720A/( ), and Agilent, Model 3458A.
(1) Connect equipment as shown in figure 3. Set multimeter Guard switch
to Open.