TB 9-4931-537-24
(3) On Calibrator, press WBAND. Set Calibrator for 1 V at first wideband output
frequency listed in table 1. At each frequency, adjust the calibrator output adjustment
control knob for a calibrator display reading equal to multimeter indication. The calibrator
control ERROR display indication will be within specified limits of table 1. Record
calibrator ERROR display indication in space provided.
(4) Set calibrator to STANDBY, then press RESET pushbutton.
Calibrator, Fluke, Model 5720A/( )
SN:____________________
Team:
_________________________________________
Team Chief Name:
Amplifier, Fluke, Model 5725A/( )
SN:____________________
_________________________________________
Signature:
SN:____________________
Calibrator
Limits (V)
Multimeter Indications
Output
Min
Max
100 mV dc
099.99730 mV
100.00270 mV
1.0 V dc
0.99998560
1.00001440
10.0 V dc
09.9998670
10.0001330
100 V dc
099.998410
100.001590
1000 V dc
0999.97495
1000.02505
10 mV, 1 kHz
09.97210 mV
10.02790 mV
100 mV, 10 kHz
099.9634 mV
100.0366 mV
1.0 V, 50 kHz
0.998093
1.001907
10 V, 20 Hz
09.98287
10.01713
10 V, 20 kHz
09.99686
10.00314
100 V, 100 Hz
099.9588
100.0412
500 V, 500 Hz
0499.498
500.502
500 V, 10 kHz1
0499.492
500.508
100 A dc
099.97900 A
100.02100 A
100 mA dc
099.99318 mA
100.00682 mA
1.0 A dc
0.9998228 A
1.0001772 A
1.0 A dc 1, 2, 3
0.9991507 A
1.0008493 A
100 A, 40 Hz
099.8240 A
100.1760 A
100 mA, 5 kHz
099.8338 mA
100.1662 mA
1.0 A, 500 Hz
0.998563 A
1.001437 A
1.0 A, 500 Hz 1,2
0.9986403 A
1.0013598 A
See footnotes at end of table.
9