TB 9-4935-552-50-2
g. Set the
differential voltmeter POWER to BAT
OPR.
h. Set and adjust radiometer A2 controls as follows:
VIEW/DETECT to VIEW.
CAL/OP to OP.
MODE to position 1.
RANGE to position 5.
NOTE
Ensure filter wheels are properly seated in detents.
Position the filter selector controls as follows:
NOTE
Auxilliary filter selector control (red markings) is permanently
locked in position.
(1) Spectral filter consul (yellow markings) to 7-69 position.
(2) Field of view (FOV) control (white markings) to 2MR position.
Focus objective lens to infinity position.
i. Plug system power supply A5 directly into 120V/400Hz outlet J3 on the test station.
Energize power supply A5.
j. Set B7 remote control POWER and PRISM switches to OFF. Connect extender
cable B7 to the TI.
k. Connect TI extender cable B7 directly to J1 in the optics bench connector well.
l. Press POWER ON switch on the test station monitor panel and install PB-109
(B10).
m. Connect one CA-135 (B5) as follows: P2 to test adapter J4, P1 to lamp driver J9.
n. Set monitor panel TEST MODE switch to MANUAL and press START TEST switch.
o. Set source/detector adapter power supply circuit breakers to ON in the following
sequence: 28 VDC, 208 VAC, LAMP DRIVER.
p. Program: SRSA00002@SPS2028@A.
7