TB 9-4935-552-50-2
Paragraph
Page
CALIBRATION PROCEDURES: IR PROBE
Radiation level ....................................................
Filter checks .......................................................
IR probe ..............................................................
Final procedure...................................................
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification
a. This bulletin provides instructions for the
calibration of
intensity test source (ITS)
11154040 (TA114) and IR probe 10684700 (TA-437). Both items are portable test
equipment in the test station of the Land Combat Support System (LCSS). Throughout the
remainder of this bulletin, the item being calibrated will be referred to as the test
instrument (TI).
b. There is only one model of each of these TIs.
c. The time required for this calibration is approximately 3 hours using the IR
technique.
2. Calibration Data Card. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25. DA Form 2416 (Calibration Data
Card) must be annotated for each calibration performed.
Test instrument parameters
and
performance
3. Calibration Description.
specifications which pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument parameters
Performance specifications
INTENSITY TEST SOURCE
Filament image focus
Edges definable to within 0.5 milliradians
Filament image intensity
25 4 watts peak-to-peak steradian
Boresight alignment
Image centered to within 4 milliradians
IR PROBE
Irradiance level
1.0x10-10 to 5x10-7 WATTS/CM2
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required
a. Table 2
identifies the
specific equipment used in
this calibration procedure. This
equipment, excluding LCSS equipment, is issued with secondary transfer calibration set
AN/GSM-286.
2