TB 9-4935-552-50-2
k. On the
positioning table, release the
azimuth and
elevation brake levers and
press
the POWER ON/OFF switch to ON.
Program: SPS5028PGNRPGNAPGNP00625P GNW003125PGN + 05PGNSRSA1@A.
l.
m. Rotate radiometer A2 optical head reflex sight lever to VIEW position.
Adjust
reticle focus as required to produce a sharp image of reticle crosshairs.
n. Loosen focus lens control on the barrel of radiometer A2 and adjust the lens barrel
in and out to produce a sharp, parallax-free image of the source/detector adapter filament.
o. Center detector adapter image in radiometer A2 optical head reticle using
positioning table controls.
p. Set and adjust radiometer A2 controls as follows:
CAL/OP to OP.
MODE to position 3.
RANGE to position 3.
NOTE
Ensure filter wheels are properly seated in detents.
Position the filter selector controls as follows:
NOTE
Auxiliary filter selector control (red markings) is permanently
locked in position.
(1) Spectral filter control (yellow markings) to 7-69 position.
(2) FOV control (white markings) to 5MR position.
q. Plug A5 directly into the 120V/400Hz outlet on the test station and energize A5.
r. Allow a 30 minute
equipment warmup
before performing the IR
probe calibration.
SECTION VI
CALIBRATION PROCEDURES: IR PROBE
a. Program SMAC11@K.
b. Connect AC/DC probe to self-test reference 1A7 (7 VAC left/right).
14