TB 9-4935-552-50-2
a. TRD shall read between 0.77 and 0.96 VAC. Record reading. No adjustment can be
made. If out of tolerance, notify LCSS maintenance personnel that the ND 0.477 filter is
faulty.
b. Program: (Press RESET then ADVANCE switches) SRSA4@K.
c. Set radiometer A2 RANGE to position 3.
d. TRD shall read between 2.33 and 2.88 VAC. Record reading. No adjustment can be
made. If out of tolerance, notify LCSS maintenance personnel that the ND 1.000 filter is
faulty.
e. Program: (Press RESET then ADVANCE switches) SRSA01@K.
f. TRD shall read between 0.77 and 0.96 VAC. Record reading. No adjustment can be
made. If out of tolerance, notify LCSS maintenance personnel that the ND 1.477 filter is
faulty.
g. Program: (Press RESET then ADVANCE switches) SRSA02@K.
h. Set radiometer A2 RANGE to position 2.
i. TRD shall read between 2.20 and 2.88 VAC. Record reading. No adjustment can be
made. If out of tolerance, notify LCSS maintenance personnel that the ND 2.000 filter is
faulty.
Program: (Press RESET then ADVANCE switches) SRSA04@K.
j.
k. TRD shall read between 0.73 and 0.96 VAC. Record reading. No adjustment can be
made. If out of tolerance, notify LCSS maintenance personnel that the ND 2.477 filter is
faulty.
Program: (Press RESET then ADVANCE switches) SRSA001@K.
l.
m. Set radiometer A2 RANGE to position 1.
n. TRD shall read between 2.12 and 2.88 VAC. Record reading. No adjustment can be
made. If out of tolerance, notify LCSS maintenance personnel that the ND 3.000 filter is
faulty.
o. Remove radiometer A2 from TA-237.
17