Calibration Procedure For Torque Wrenches, Powerdyne, Models Pd 1220, Pd 1220B, Pd 1201, And Pd 2501; Consolidated Devices, Models Dpt-1200, Dpt-1200R, Dpt-2500, And Dpt-2500R; Torque Applicator, Powerdyne Model Pd 12003 And Advanced Torque Models Ctw 1200-T55 And Atp 8000 Page Navigation
  15    16    17    18    19  20
img
. . .
TB 9-5120-212-24
NOTE
In order to set the anti-wind-up knob from CW to CCW a small
amount of cw force will need to be applied.
(b) Set anti-wind-up knob on TI to CCW. Slowly release pull handle cw to
obtain a 0 ft-lbs reading on the force/torque indicator.
(c) Repeat (a) and (b) above two times.
(11) The TI display will prompt "FULL LOAD CW", turn pull handle cw to obtain a
reading of 7000 ft-lbs on the force/torque indicator.
(12) After the load has stabilized press the RESET pad.
(13) The TI display will prompt "IF YOU ARE SURE PRESS RESET", press the
RESET pad to enter the calibration values. Remove load from TI.
(14) Repeat paragraph 20.
21. Final Procedure
a. Deenergize and disconnect all equipment.
b. For model ATP 8000 affix DA Label 163 (checked to 2% FS due to limitations of
standards).
b. Annotate and affix DA label/form in accordance with TB 750-25.
20