TB 9-5120-212-24
CALIBRATION PROCESS FOR
ADVANCED TORQUE, MODEL ATP 8000
18. Preliminary Instructions
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
NOTE
View the advisory by clicking the *** (triple asterisk) in the cal
responsibility column of the 43-180 on the USATA home page.
a. Verify that TI is clean and free from defects that would impair its operation.
b. Ensure that the equipment has been allowed to stabilize at the ambient
temperature.
c. Position torque plate on a stable and rigid work surface and secure with bolts or
clamps.
d. Assemble TI and torque cell no. 2 on torque plate using appropriate drive bar and
socket wrench adapter. Ensure that locating pins and drive bar are engaged in their
respective mounting holes. Install TI crank handle in socket as necessary.
e. Connect torque cell to force/torque indicator using cable supplied with force/torque
indicator. Connect force/torque indicator to appropriate power source. Turn power switch
to ON and allow the units to warm-up for 30 minutes. Select channel 3 using the
CHANNEL + and keys.
f. Press the F4 key on the force/torque indicator until the
soft keys are
displayed. Press the
soft keys as necessary to select the torque cell for the TI range
and ccw torque. Ensure the serial number matches the torque cell being used.
NOTE
Turning the TI crank cw will cause ccw torque on the torque cell.
g. Press the F4 key on the force/torque indicator until the UNIT soft key is displayed.
Press the UNIT soft key as necessary to select ftlb.