TB 9-5210-207-24
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance applications which pertain
to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
1/ to 42 in.
Length (English)
Range: 1 2
0.001 in. (without vernier scale)
Accuracy:
0.0001 in. (with vernier scale)
Length (Metric)
Range: 25 to 800 mm
0.01 mm
Accuracy:
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, AN/GSM-705 and Secondary Reference
Calibration Standards Set NSN 4931-00-621-7878. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a
four-to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met,
the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.